
Analyzers and Collision Cell
7
Troubleshooting and Maintenance Manual
147
The collision cell
The precursor ions selected by the front quadrupole collide with a reaction
gas, typically a nitrogen/helium mixture. This generates additional product
ions.
The pre-filter
This device, located on the entrance of the rear analyzer, transfers the
remaining precursor ions and the product ions generated in the collision cell
to the rear analyzer. It also aligns the rear of the collision cell with the rear
analyzer.
The rear analyzer
The rear analyzer consists of the rear quadrupole, the detector, and a
heater/sensor assembly.
The ion stream now consists of the ions selected by the front analyzer and the
ions formed by collision with the reaction gas. The rear analyzer separates
these ions and passes them to the detector.
The detector
Ions that pass the rear analyzer strike a high-energy dynode, where they
generate secondary electrons. These enter a multiplier and are collected to
become the output signal.
Summary of Contents for 7000 Triple Quadrupole GC/MS
Page 24: ...24 Troubleshooting and Maintenance Manual 1 Introduction...
Page 50: ...50 Troubleshooting and Maintenance Manual 2 General Troubleshooting...
Page 122: ...122 Troubleshooting and Maintenance Manual 4 General Maintenance...
Page 144: ...144 Troubleshooting and Maintenance Manual 6 Vacuum System...
Page 182: ...182 Troubleshooting and Maintenance Manual 8 Electronics...
Page 213: ......
Page 214: ...Agilent Technologies Agilent Technologies Inc Printed in USA September 2009...