62
Troubleshooting and Maintenance Manual
3
CI Troubleshooting
Signal-Related Symptoms
This section describes symptoms related to the signal. The symptom may be
too much signal, too little signal, a noisy signal, or an incorrect signal.
Signal-related symptoms are generally observed during tuning but may also be
observed during data acquisition.
Error messages in autotune due to insufficient signal may vary.
The following symptoms are covered in more detail in this section:
•
.
•
No or low reagent gas signal. See
•
No or low PFDTD signal. See
.
•
Excessive noise. See
•
Low signal-to-noise ratio. See
•
Large peak at
m/z
19. See
.
•
Peak at
m/z
.
No peaks
When troubleshooting “no peaks” it is important to specify what mode of
operation is being used and what expected peaks are not being seen. Always
start with methane PCI and verify presence of reagent ions.
No reagent gas peaks in PCI
If MS has been working well and nothing seems to have been changed
•
Wrong tune file loaded, or tune file corrupted
•
Wrong ion polarity (there are no reagent ions visible in NCI)
•
No reagent gas flow; look for background ions and check pressure
•
Wrong reagent gas selected for the tune file (looking for wrong ions)
•
Large air leak
•
Dirty ion source
•
Poor vacuum (pump problem). See
.
Summary of Contents for 7000 Triple Quadrupole GC/MS
Page 24: ...24 Troubleshooting and Maintenance Manual 1 Introduction...
Page 50: ...50 Troubleshooting and Maintenance Manual 2 General Troubleshooting...
Page 122: ...122 Troubleshooting and Maintenance Manual 4 General Maintenance...
Page 144: ...144 Troubleshooting and Maintenance Manual 6 Vacuum System...
Page 182: ...182 Troubleshooting and Maintenance Manual 8 Electronics...
Page 213: ......
Page 214: ...Agilent Technologies Agilent Technologies Inc Printed in USA September 2009...