32
Troubleshooting and Maintenance Manual
2
General Troubleshooting
Baseline wanders
•
Insufficient carrier gas supply pressure*
•
Malfunctioning flow or pressure regulator*
•
Intermittent leak in the GC inlet*
* This could cause a fault condition in the GC that would prevent the GC
from operating.
Retention times for all peaks drift – shorter
•
Column has been shortened
•
Initial oven temperature was increased
•
Column is getting old
Retention times for all peaks drift – longer
•
Column flow has been reduced
•
Initial oven temperature was decreased
•
Active sites in the sample path
•
Leaks in the GC inlet*
* This could cause a fault condition in the GC that would prevent the GC
from operating.
Poor sensitivity
•
Incorrect tuning, or tune file that does not match the type of analysis
•
Repeller voltage is too low
•
Incorrect temperatures (oven, GC/MS interface, ion source, or mass filter)
•
Incorrect sample concentration
•
Leaking GC inlet*
•
Dirty GC inlet
•
Incorrect split ratio
•
Purge-off time in splitless mode is too short