
General Troubleshooting
2
Troubleshooting and Maintenance Manual
49
Table 3
Common contaminants
Ions (
m/z
)
Compound
Possible source
18, 28, 32, 44 or 14, 16
H
2
O, N
2
, O
2
, CO
2
or N, O
Residual air and water, air
leaks, outgassing from Vespel
ferrules
31, 51, 69, 100, 119, 131,
169, 181, 214, 219, 264, 376,
414, 426, 464, 502, 576, 614
PFTBA and related ions
PFTBA (tuning compound)
31
Methanol
Cleaning solvent
43, 58
Acetone
Cleaning solvent
78
Benzene
Cleaning solvent
91, 92
Toluene or xylene
Cleaning solvent
105, 106
Xylene
Cleaning solvent
151, 153
Trichloroethane
Cleaning solvent
69
Foreline pump oil or PFTBA
Foreline pump oil vapor or
calibration valve leak
73, 147, 207, 221, 281, 295, 355,
429
Dimethylpolysiloxane
Septum bleed or methyl
silicone column bleed
149
Plasticizer (phthalates)
Vacuum seals (O-rings)
damaged by high
temperatures, vinyl gloves
Peaks spaced 14
m/z
apart
Hydrocarbons
Fingerprints, foreline pump oil
Summary of Contents for 7000 Triple Quadrupole GC/MS
Page 24: ...24 Troubleshooting and Maintenance Manual 1 Introduction...
Page 50: ...50 Troubleshooting and Maintenance Manual 2 General Troubleshooting...
Page 122: ...122 Troubleshooting and Maintenance Manual 4 General Maintenance...
Page 144: ...144 Troubleshooting and Maintenance Manual 6 Vacuum System...
Page 182: ...182 Troubleshooting and Maintenance Manual 8 Electronics...
Page 213: ......
Page 214: ...Agilent Technologies Agilent Technologies Inc Printed in USA September 2009...