
General Troubleshooting
2
Troubleshooting and Maintenance Manual
47
Air Leaks
Air leaks are a problem for any instrument that requires a vacuum to operate.
Leaks are generally caused by vacuum seals that are damaged or not fastened
correctly. Symptoms of leaks include:
•
Higher than normal analyzer chamber pressure or foreline pressure
•
Higher than normal background
•
Peaks characteristic of air (
m/z
18, 28, 32, and 44 or
m/z
14 and 16)
•
Poor sensitivity
•
Low relative abundance of
m/z
502 (this varies with the tune program used)
Leaks can occur in either the GC or the MS. The most likely point for an air
leak is a seal you recently opened.
In the GC, most leaks occur in:
•
GC inlet septum
•
GC inlet column nut
•
Broken or cracked capillary column
Leaks can occur in many more places in the MS:
•
GC/MS interface column nut
•
Side plate O-rings (all the way around)
•
Vent valve O-ring
•
Calibration valve
•
GC/MS interface O-ring (where the interface attaches to the analyzer
chamber)
•
Front and rear end plate O-rings
•
Turbo pump O-rings
•
Collision cell cover O-ring
Summary of Contents for 7000 Triple Quadrupole GC/MS
Page 24: ...24 Troubleshooting and Maintenance Manual 1 Introduction...
Page 50: ...50 Troubleshooting and Maintenance Manual 2 General Troubleshooting...
Page 122: ...122 Troubleshooting and Maintenance Manual 4 General Maintenance...
Page 144: ...144 Troubleshooting and Maintenance Manual 6 Vacuum System...
Page 182: ...182 Troubleshooting and Maintenance Manual 8 Electronics...
Page 213: ......
Page 214: ...Agilent Technologies Agilent Technologies Inc Printed in USA September 2009...