110
Troubleshooting and Maintenance Manual
4
General Maintenance
5
Separate the source heater assembly from the source body by removing
the two screws. The source heater assembly includes the source heater,
repeller, and related parts. (See
.)
6
Disassemble the repeller assembly by removing the repeller nut, washers,
ceramic insulators and repeller. (See
7
Unscrew the interface socket from the source body. A 10 mm open-end
wrench fits the flat sides on the interface socket.
C A U T I O N
When disconnecting leads, pull on the connectors, not on the wires.
Figure 13
Replacing the heater and sensor
Filament
Filament
Repeller assembly
Repeller nut
Source radiator