Troubleshooting and Maintenance Manual
2
General Troubleshooting
There is no emission current
•
Check tune file to be certain that emission current is not = 0
•
Filament is not connected properly; try the other filament
•
Filament has failed; try the other filament
•
MS electronics are not working correctly
There is not enough signal to begin tune
•
Corrupted tune file
•
Poor mass axis calibration
•
Width gain or offset is too high
•
Calibration vial(s) empty or almost empty
•
Excessive pressure in the analyzer chamber
•
Air leak
•
Electron multiplier voltage is too low
•
Signal cable is not connected
•
Electrical leads to the detector are not connected correctly
•
Electrical leads to the ion source are not connected correctly
•
Filament shorted to the source body
•
Collision cell gas flow
•
Collision cell voltages