152
Troubleshooting and Maintenance Manual
7
Analyzers and Collision Cell
Filaments
Two filaments (
and
) are located on opposite sides of the
outside of the EI ion source. The
active
filament carries an adjustable AC
emission current. The emission current heats the filament causing it to emit
electrons which ionize the sample molecules. In addition,
both
filaments have
an adjustable DC bias voltage. The bias voltage determines the energy on the
electrons, usually –70 eV.
The CI ion source has only one filament of a different design from the EI
filaments. A “dummy” filament provides connections for the Filament 2 wires.
The filament is shut off automatically if there is a general instrument
shutdown. Three parameters affect the filaments: filament selection
(Filament), filament emission (Emission) current, and electron energy
(EIEnrgy).
Filament selection
The filament selection parameter (Filament) selects which filament in the ion
source is active. In the CI source, it is always Filament 1.
Since filaments may have some variability, we advise running a tune for each
filament and saving the tune accordingly.
Emission current
The filament emission current (Emission) is variable between 0 and –315 µA,
but should be set to the software default for normal operation.
Electron energy
The electron energy (EIEnrgy) is the amount of energy on the ionizing
electrons. It is determined by the bias voltage; –70 VDC bias on the filament
causes emitted electrons to possess –70 eV (electron volts). This value is
adjustable from –5 to –241 VDC, but for normal operation, set this parameter
to 70.
Summary of Contents for 7000 Triple Quadrupole GC/MS
Page 24: ...24 Troubleshooting and Maintenance Manual 1 Introduction...
Page 50: ...50 Troubleshooting and Maintenance Manual 2 General Troubleshooting...
Page 122: ...122 Troubleshooting and Maintenance Manual 4 General Maintenance...
Page 144: ...144 Troubleshooting and Maintenance Manual 6 Vacuum System...
Page 182: ...182 Troubleshooting and Maintenance Manual 8 Electronics...
Page 213: ......
Page 214: ...Agilent Technologies Agilent Technologies Inc Printed in USA September 2009...