112
Troubleshooting and Maintenance Manual
4
General Maintenance
To remove/reinstall the heater and sensor for the mass filter
Materials needed
•
Gloves, clean, lint-free
•
Large (8650-0030)
•
Small (8650-0029)
•
Hex ball driver, 1.5 mm (8710-1570)
•
Hex ball driver, 2.0 mm (8710-1804)
Removal
1
Vent the MS. See the
Agilent 7000 Series Triple Quad GC/MS Operation
.
2
Open the front or rear analyzer chamber.
3
Disconnect the mass filter heater and temperature sensor wires from the
feedthrough board.
4
Remove the mass filter heater assembly from the mass filter radiator by
removing the two screws. See
for the front mass filter. The rear
mass filter heater/sensor removal is similar.
WA R N I N G
The analyzers, GC/MS interface, and other components in the analyzer chamber
operate at very high temperatures. Do not touch any part until you are sure it is cool.
C A U T I O N
Always wear clean gloves to prevent contamination when working in the analyzer
chamber.
C A U T I O N
Make sure you use an antistatic wrist strap and take other antistatic precautions
before touching analyzer components.
C A U T I O N
When disconnecting leads, pull on the connectors, not on the wires.
Summary of Contents for 7000 Triple Quadrupole GC/MS
Page 24: ...24 Troubleshooting and Maintenance Manual 1 Introduction...
Page 50: ...50 Troubleshooting and Maintenance Manual 2 General Troubleshooting...
Page 122: ...122 Troubleshooting and Maintenance Manual 4 General Maintenance...
Page 144: ...144 Troubleshooting and Maintenance Manual 6 Vacuum System...
Page 182: ...182 Troubleshooting and Maintenance Manual 8 Electronics...
Page 213: ......
Page 214: ...Agilent Technologies Agilent Technologies Inc Printed in USA September 2009...