General Maintenance
4
Troubleshooting and Maintenance Manual
95
Removal, analyzer chamber gauge
1
Vent the MS. See the
Agilent 7000 Series Triple Quad GC/MS Operation
.
2
The chamber ion gauge is under the collision cell area. See
.
3
Separate the MS from the GC. See
“To separate/connect the MS and the
4
Remove the cable from the bottom of the ion gauge.
5
Remove the screws from around the base of the ion gauge.
6
Remove the gauge and O-ring from the chamber.
Installation, analyzer chamber gauge
Installation is the reverse of the removal procedure.
Removal, foreline pump gauge
1
Vent the MS. See the
Agilent 7000 Series Triple Quad GC/MS Operation
.
2
Separate the MS from the GC. See
“To separate/connect the MS and the
3
Unplug the foreline gauge cable from the foreline gauge.
4
Unscrew the large wingnut on the gauge clamp. See
5
While supporting the gauge body, remove the clamp from the mounting
flange.
6
Remove the gauge and hose together.
7
Loosen the hose clamp.
8
Pull the foreline gauge assembly out of the foreline hose.
Installation, foreline pump gauge
Installation is the reverse of the removal procedure.
Summary of Contents for 7000 Triple Quadrupole GC/MS
Page 24: ...24 Troubleshooting and Maintenance Manual 1 Introduction...
Page 50: ...50 Troubleshooting and Maintenance Manual 2 General Troubleshooting...
Page 122: ...122 Troubleshooting and Maintenance Manual 4 General Maintenance...
Page 144: ...144 Troubleshooting and Maintenance Manual 6 Vacuum System...
Page 182: ...182 Troubleshooting and Maintenance Manual 8 Electronics...
Page 213: ......
Page 214: ...Agilent Technologies Agilent Technologies Inc Printed in USA September 2009...