CI Troubleshooting
3
Troubleshooting and Maintenance Manual
63
If MS was recently switched from EI to CI
•
Interface tip seal not installed
•
No reagent gas flow
•
Analyzer not sealed (big air leak)
•
Wrong tune file loaded or tune file corrupted
•
Ion source not assembled or connected correctly
•
Wrong reagent gas selected for the tune file (looking for wrong ions)
No PFDTD peaks in PCI
•
Incorrect reagent gas. There
are
no PCI PFDTD peaks created with
isobutane or ammonia. Switch to methane.
•
Analyzer not sealed (big air leak)
•
No calibrant in vial
•
Defective calibration valve(s)
•
Air leak in carrier or reagent gas path
No reagent gas peaks in NCI
•
Reagent gases do not ionize in NCI; look for background ions instead
•
Verify tune parameters
•
If no background ions are visible, go back to methane PCI
No PFDTD calibrant peaks in NCI
•
Look for background ions: 17 (OH
–
), 35 (Cl
–
), and 235 (ReO3
–
)
•
Verify tune parameters
•
Go back to methane PCI
No sample peaks in NCI
•
Look for background ions: 17 (OH
–
), 35 (Cl
–
), and 235 (ReO3
–
)
•
Go back to methane PCI
•
Poor quality reagent gas (purity less than 99.99%)
Large peak at
m/z
238 in NCI OFN spectrum
•
Look for background ions: 17 (OH
–
), 35 (Cl
–
), and 235 (ReO3
–
)
•
Find and fix your small air leak
Summary of Contents for 7000 Triple Quadrupole GC/MS
Page 24: ...24 Troubleshooting and Maintenance Manual 1 Introduction...
Page 50: ...50 Troubleshooting and Maintenance Manual 2 General Troubleshooting...
Page 122: ...122 Troubleshooting and Maintenance Manual 4 General Maintenance...
Page 144: ...144 Troubleshooting and Maintenance Manual 6 Vacuum System...
Page 182: ...182 Troubleshooting and Maintenance Manual 8 Electronics...
Page 213: ......
Page 214: ...Agilent Technologies Agilent Technologies Inc Printed in USA September 2009...