Reinstallation
Materials needed
•
GC/MS interface heater assembly (G1099-60107)
•
Heater clamp (G1099-20210); replace the old clamp if it is damaged
•
Screwdriver, Torx T-15 (8710-1622)
•
Hex driver, 1.5 mm (8710-1570)
C A U T I O N
Installing a new heater and sensor in a damaged heater sleeve will result in poor
performance of the heated zone and could reduce the lifetime of the new parts.
Figure 16
Replacing the GC/MS interface heater and sensor
Heater clamp
Setscrew
Temperature sensor
Heater
Insulation
Cover
MS
Tip of interface