58
Troubleshooting and Maintenance Manual
3
CI Troubleshooting
Pressure-Related Symptoms
The following symptoms are all related to high vacuum pressure. Each
symptom is discussed in more detail in the following pages.
The mass flow controller is calibrated for methane and the high vacuum gauge
controller is calibrated for nitrogen, so these measurements are not accurate
in absolute terms (
). They are intended as a guide to typical observed
readings. They were taken with the following set of conditions:
Source temperature
300 °C
Quad temperature
150 °C
Interface temperature
280 °C to 320 °C
Helium carrier gas flow
1 mL/min
Table 5
Typical analyzer vacuum with reagent gas flow
Collision cell gas flow on
Collision cell gas flow off
MFC
(%)
Rough vac
High vac
Rough vac
High vac
10
1.77
×
10
–1
7.15
×
10
–5
1.33
×
10
–1
2.56
×
10
–6
15
1.86
×
10
–1
7.19
×
10
–5
1.43
×
10
–1
3.00
×
10
–6
20
1.94
×
10
–1
7.23
×
10
–5
1.53
×
10
–1
3.45
×
10
–6
25
2.02
×
10
–1
7.27
×
10
–5
1.63
×
10
–1
3.86
×
10
–6
30
2.10
×
1
0
–1
7.31
×
10
–5
1.71
×
1
0
–1
4.30
×
10
–6
35
2.18
×
10
–1
7.39
×
10
–5
1.80
×
10
–1
4.76
×
10
–6
40
2.25
×
10
–1
7.43
×
10
–5
1.88
×
10
–1
5.18
×
10
–6
Summary of Contents for 7000 Triple Quadrupole GC/MS
Page 24: ...24 Troubleshooting and Maintenance Manual 1 Introduction...
Page 50: ...50 Troubleshooting and Maintenance Manual 2 General Troubleshooting...
Page 122: ...122 Troubleshooting and Maintenance Manual 4 General Maintenance...
Page 144: ...144 Troubleshooting and Maintenance Manual 6 Vacuum System...
Page 182: ...182 Troubleshooting and Maintenance Manual 8 Electronics...
Page 213: ......
Page 214: ...Agilent Technologies Agilent Technologies Inc Printed in USA September 2009...