General Maintenance
4
Troubleshooting and Maintenance Manual
103
To lubricate the side plate O-rings
Materials needed
•
Cloths, clean (05980-60051)
•
Gloves, clean, lint-free
•
Large (8650-0030)
•
Small (8650-0029)
•
Grease, Apiezon L, high-vacuum (6040-0289)
The side plate O-rings may require a thin coat of grease to ensure a good
vacuum seal. If an O-ring appears dry or does not seal correctly, lubricate it
using this procedure. A good test is to wipe off the side plate with methanol,
then close the analyzer chamber. If the O-ring has enough grease on it, it will
leave a faint trace on the side plate.
Procedure
1
Vent the MS. See the
Agilent 7000 Series Triple Quad GC/MS Operation
.
2
Open an analyzer chamber.
3
Use a clean, lint-free cloth or glove to spread a
thin
coat of high-vacuum
grease only on the exposed surface of the O-ring (
WA R N I N G
The analyzer, GC/MS interface, and other components in the analyzer chamber
operate at very high temperatures. Do not touch any part until you are sure it is cool.
C A U T I O N
Vacuum seals other than the side plate O-ring and vent valve O-ring do not need to be
lubricated. Lubricating other seals can interfere with their correct function.
C A U T I O N
Always wear clean gloves to prevent contamination when working in the analyzer
chamber.
C A U T I O N
Electrostatic discharges to analyzer components are conducted to the quad driver
board, where they can damage sensitive components. Wear a grounded antistatic
wrist strap and take other antistatic precautions (see
on page 13) before you open the analyzer chamber.
Summary of Contents for 7000 Triple Quadrupole GC/MS
Page 24: ...24 Troubleshooting and Maintenance Manual 1 Introduction...
Page 50: ...50 Troubleshooting and Maintenance Manual 2 General Troubleshooting...
Page 122: ...122 Troubleshooting and Maintenance Manual 4 General Maintenance...
Page 144: ...144 Troubleshooting and Maintenance Manual 6 Vacuum System...
Page 182: ...182 Troubleshooting and Maintenance Manual 8 Electronics...
Page 213: ......
Page 214: ...Agilent Technologies Agilent Technologies Inc Printed in USA September 2009...