92
Troubleshooting and Maintenance Manual
4
General Maintenance
Reinstallation
Materials needed
•
Calibration valve (G3170-60204)
•
O-ring for calibration valve (0905-1217)
•
PFTBA (05971-60571)
•
Screwdriver, Torx T-15 (8710-1622)
1
Remove the old valve assembly. See
“To remove/reinstall the calibration
2
Be sure the valve O-ring is in place. If it is worn or damaged, replace it.
3
Install the calibration and vent valve assembly and tighten the screws that
hold it in place.
4
Reconnect the calibration valve cable to the connector.
5
Remove the vial from the new calibration valve. See
on page 89. The valve is supplied with a vial already
installed.
6
Fill and reinstall the calibration vial. See
7
Pump down the MS. See the
Agilent 7000 Series Triple Quad GC/MS
C A U T I O N
Do not turn on any GC heated zones until the carrier gas flow is on. Heating a column
with no carrier gas flow will damage the column.