General Troubleshooting
Temperature Symptoms
The MS has four heated zones:
•
Ion source
•
Front and rear mass filters
•
GC/MS interface
Each heated zone has a heater and temperature sensor. The ion source and
mass filters are powered and controlled by the MS. The GC/MS interface is
powered and controlled by the GC.
Ion source will not heat up
•
High-vacuum pump is off or has not reached normal operating conditions*
•
Incorrect temperature setpoint
•
Ion source has not had enough time to reach temperature setpoint
•
Ion source heater cartridge is not connected*
•
Ion source temperature sensor is not connected*
•
Ion source heater failed (burned out or shorted to ground)*
•
Ion source temperature sensor failed*
•
Source power cable is not connected to the quadrupole board*
•
MS electronics are not working correctly
* This will cause an error message.
Mass filter (quad) heaters will not heat up
•
High-vacuum pump is off or has not reached normal operating conditions*
•
Incorrect temperature setpoint
•
Mass filter has not had enough time to reach temperature setpoint
•
Mass filter heater cartridge is not connected*
•
Mass filter temperature sensor is not connected*
•
Mass filter heater failed (burned out or shorted to ground)*
•
Mass filter temperature sensor failed*
•
Cable is not connected to the quadrupole board*
Summary of Contents for 7000 Triple Quadrupole GC/MS
Page 24: ...24 Troubleshooting and Maintenance Manual 1 Introduction...
Page 50: ...50 Troubleshooting and Maintenance Manual 2 General Troubleshooting...
Page 122: ...122 Troubleshooting and Maintenance Manual 4 General Maintenance...
Page 144: ...144 Troubleshooting and Maintenance Manual 6 Vacuum System...
Page 182: ...182 Troubleshooting and Maintenance Manual 8 Electronics...
Page 213: ......
Page 214: ...Agilent Technologies Agilent Technologies Inc Printed in USA September 2009...