CI Troubleshooting
3
Troubleshooting and Maintenance Manual
65
The signal cable is not connected.
Check and, if necessary, reconnect the signal cable.
The filament or filament support is shorted to the ion source body or repeller.
Inspect the filament. If necessary, realign the filament support arms.
The electron inlet hole is blocked.
Inspect the electron inlet hole. If necessary, clean the hole with a clean
toothpick and a slurry of aluminum oxide powder and methanol. If the
electron inlet hole is that dirty, the entire ion source probably needs to be
cleaned. See
on page 75 for more
information.
Ion source wires are not connected, or incorrectly connected.
Inspect the repeller. Make sure the repeller lead is firmly attached to the
repeller. Inspect the wires to the ion focus and entrance lenses. If the
connections are reversed, correct the problem.
One of the detector leads (in the analyzer chamber) is not connected.
Check and, if necessary, reconnect the electron multiplier leads.
Saturated methane/isobutane gas purifier
Replace the gas purifier.
Poor quality methane (purity below 99.99%)
Replace the methane with high-purity methane. If necessary, clean and purge
the reagent gas lines and clean the ion source.
Summary of Contents for 7000 Triple Quadrupole GC/MS
Page 24: ...24 Troubleshooting and Maintenance Manual 1 Introduction...
Page 50: ...50 Troubleshooting and Maintenance Manual 2 General Troubleshooting...
Page 122: ...122 Troubleshooting and Maintenance Manual 4 General Maintenance...
Page 144: ...144 Troubleshooting and Maintenance Manual 6 Vacuum System...
Page 182: ...182 Troubleshooting and Maintenance Manual 8 Electronics...
Page 213: ......
Page 214: ...Agilent Technologies Agilent Technologies Inc Printed in USA September 2009...