General Maintenance
4
Troubleshooting and Maintenance Manual
115
Maintaining the GC/MS Interface
The GC/MS interface requires no periodic maintenance
Rarely, the heater cartridge in the GC/MS interface fails. In those cases, it is
necessary to replace the heater and sensor. This section contains procedures
for removing the heater and sensor and installing new ones. See
.
Figure 15
GC/MS interface
MS
GC oven
Ionization chamber
Heater/sensor assembly
Analyzer chamber
Heater sleeve
Insulation