General Troubleshooting
2
Troubleshooting and Maintenance Manual
29
Chromatographic Symptoms
These are symptoms you may observe in the chromatograms generated by data
acquisition. In general, these symptoms do not prevent you from operating
your GC/MS system. They indicate, however, that the data you are acquiring
may not be the best data obtainable. These symptoms can be caused by
instrument malfunctions but are more likely caused by incorrect
chromatographic technique.
Two of the symptoms,
Low sensitivity
and
Poor repeatability
also apply to
mass spectral data.
No peaks
If an analysis shows no chromatographic peaks, only a flat baseline or minor
noise, run the automated tune program. If the MS passes tune, the problem is
most likely related to the GC. If the MS does not pass tune, the problem is most
likely in the MS.
Passes tune
•
Incorrect sample concentration
•
No analytes present
•
Syringe missing from the ALS or not installed correctly
•
Injection accidentally made in split mode instead of splitless mode
•
Empty or almost empty sample vial
•
Dirty GC inlet
•
Leaking GC inlet*
•
Loose column nut at the GC inlet*
* This could cause a fault condition in the GC that would prevent the GC
from operating.
Does not pass tune
•
Calibration vial is empty
•
Excessive foreline or analyzer chamber pressure
•
Very dirty ion source
Summary of Contents for 7000 Triple Quadrupole GC/MS
Page 24: ...24 Troubleshooting and Maintenance Manual 1 Introduction...
Page 50: ...50 Troubleshooting and Maintenance Manual 2 General Troubleshooting...
Page 122: ...122 Troubleshooting and Maintenance Manual 4 General Maintenance...
Page 144: ...144 Troubleshooting and Maintenance Manual 6 Vacuum System...
Page 182: ...182 Troubleshooting and Maintenance Manual 8 Electronics...
Page 213: ......
Page 214: ...Agilent Technologies Agilent Technologies Inc Printed in USA September 2009...