General Troubleshooting
2
Troubleshooting and Maintenance Manual
31
Peaks have split tops
•
Bad injection technique
•
Injection is too large
Baseline is rising
•
Column bleed
•
Other contamination
Baseline is high
•
Column bleed
•
Other contamination
•
Electron multiplier voltage is too high
Baseline is falling
A falling baseline indicates contamination is being swept away. Wait until the
baseline reaches an acceptable level. Common causes include:
•
Residual air and water from a recent venting
•
Column bleed
•
Septum bleed
•
Splitless injection time too long (inlet is not properly swept, resulting in
excess solvent on the column and slow solvent decay)