
Analyzers and Collision Cell
7
Troubleshooting and Maintenance Manual
155
Ion focus
The voltage on the ion focus lens (
) can be varied from 0 to
–127 VDC. A typical voltage is between –70 and –90 VDC. In general:
•
Increasing the ion focus voltage improves sensitivity at lower masses.
•
Decreasing the ion focus voltage improves sensitivity at higher masses.
•
Incorrect ion focus adjustment results in poor high mass response.
Entrance lens
The entrance lens (
) is at the entrance to the quadrupole mass filter.
This lens minimizes the fringing fields of the quadrupole which discriminate
against high-mass ions. There is a per4.4 volt voltage added to the
entrance lens. The total voltage applied to the entrance lens is the sum of the
entrance lens offset and entrance lens gain and the +4.4 volt permanent offset.
Entrance lens voltage = +4.4 VDC + (gain × mass)
Summary of Contents for 7000 Triple Quadrupole GC/MS
Page 24: ...24 Troubleshooting and Maintenance Manual 1 Introduction...
Page 50: ...50 Troubleshooting and Maintenance Manual 2 General Troubleshooting...
Page 122: ...122 Troubleshooting and Maintenance Manual 4 General Maintenance...
Page 144: ...144 Troubleshooting and Maintenance Manual 6 Vacuum System...
Page 182: ...182 Troubleshooting and Maintenance Manual 8 Electronics...
Page 213: ......
Page 214: ...Agilent Technologies Agilent Technologies Inc Printed in USA September 2009...