General Troubleshooting
2
Troubleshooting and Maintenance Manual
39
Foreline pressure is too low
If the pressures you observe are below 20 mTorr, check for the following:
•
Column (carrier gas) flow is too low
•
Column plugged or crushed by an overtightened nut
•
Collision gas flows are too low
•
Empty or insufficient carrier gas supply*
•
Bent or pinched carrier gas tubing*
•
Foreline gauge is not working correctly
* This could create a fault condition in the GC that would prevent the GC
from operating.
Analyzer chamber pressure is too low
If the pressures you observe are below 1 × 10
-6
Torr, check for the following:
•
Column (carrier gas) flow is too low
•
Collision gas flows are too low
•
Column plugged or crushed by overtightened nut
•
Empty or insufficient carrier gas supply*
•
Bent or pinched carrier gas tubing*
* This could create a fault condition in the GC that would prevent the GC
from operating.
Summary of Contents for 7000 Triple Quadrupole GC/MS
Page 24: ...24 Troubleshooting and Maintenance Manual 1 Introduction...
Page 50: ...50 Troubleshooting and Maintenance Manual 2 General Troubleshooting...
Page 122: ...122 Troubleshooting and Maintenance Manual 4 General Maintenance...
Page 144: ...144 Troubleshooting and Maintenance Manual 6 Vacuum System...
Page 182: ...182 Troubleshooting and Maintenance Manual 8 Electronics...
Page 213: ......
Page 214: ...Agilent Technologies Agilent Technologies Inc Printed in USA September 2009...