80
Troubleshooting and Maintenance Manual
4
General Maintenance
Maintaining the Mainframe
The mainframe consists of everything that does not fit in the vacuum,
analyzer, interface, or electronics categories.
To remove the MS covers
Materials needed
•
Screwdriver, Torx T-15 (8710-1622)
If you need to remove one of the MS covers (see
), follow these
procedures:
To remove the top cover
Remove the seven screws and lift the cover off.
To remove the front right analyzer window cover
1
Pull out from the top of the window and forward from the bottom. The
cover is held in place by magnets.
2
Lift the window forward and off the MS.
To remove the front bottom cover
Grasp the cover on both sides and pull the cover forward.
To open the left side panel
1
Remove the front right analyzer window.
2
Pull gently on the front left window and allow the left side panel to swing
forward and down.
WA R N I N G
The GC/MS interface, the analyzer parts, and the vacuum system operate at
temperatures high enough to cause serious burns. Give these parts enough time to
cool before accessing them or handling them.
Summary of Contents for 7000 Triple Quadrupole GC/MS
Page 24: ...24 Troubleshooting and Maintenance Manual 1 Introduction...
Page 50: ...50 Troubleshooting and Maintenance Manual 2 General Troubleshooting...
Page 122: ...122 Troubleshooting and Maintenance Manual 4 General Maintenance...
Page 144: ...144 Troubleshooting and Maintenance Manual 6 Vacuum System...
Page 182: ...182 Troubleshooting and Maintenance Manual 8 Electronics...
Page 213: ......
Page 214: ...Agilent Technologies Agilent Technologies Inc Printed in USA September 2009...