PROPRIETARY INFORMATION
CONFIDENTIAL MATERIAL
Not for Customer Distribution
Assays/Reagents
AEROSET
®
System Troubleshooting Guide
57
94816-107—November 2004
Theo
Antibody to Theophylline (R1) + G-6-PD-Theophylline (R2) (Active)
G-6-PD-Theophylline:Antibody complex (Inactive)
G-6-PD-Theophylline:Antibody complex (Inactive) + Theophylline
G-6-PD-Theophylline (Active) + Theophylline:Antibody complex
G-6-PD-Theophylline (Active) + NAD + Glucose-6-phosphate
NADH + 6-phosphogluconate
G-6-PD = Glucose-6-phosphate dehydrogenase from bacterial source
NADH = Nicotinamide adenine dinucleotide
Blank this assay with
Theo CAL 1.
When either the R1 or
R2 reagent cartridge is
empty, replace both
cartridges and
validate the system by
analyzing controls.
TIBC
Apotrans excess Fe
+3
Fe-trans Fe
+3
Fe
+3
+ Fe-trans alumina
Eluent of Fe-transferrin;
analyzed using Iron
Reagent (LN 7D68)
Serum only.
TOBIL
(LN
2K58-20,
DCL
Label)
Total (Conjugated & Unconjugated) Bil 2,4-Dichlorophenyl
diazonium salt
Azobilirubin
The formation of
azobilirubin is
measured
spectrophoto-
metrically and is
directly proportional
to the concentration
of bilirubin.
Avoid hemolysis
during sample
collection. Protect
samples from light.
Same formulation as
LN 8G62-20.
TP
Cu
+2
Cu
+2
-Protein Nitrogen complex
(Biuret reaction)
Dextran interferes
with the Total Protein
assay, a biuret
method. Plasma
values are generally
0.3 to 0.5 g/dL higher
than serum values,
due to the presence of
fibrinogen.
Assay
Name
Reaction
Tips
* MULTIGENT
®
assay
Saturating Solution
Adsorbent
Columns
Surfactant
OH
-