Hardware
AEROSET
®
System Troubleshooting Guide
229
94816-107—November 2004
Reagent Supply Centers have no water
Probable Cause(s)
Corrective Action(s)
Water evaporated.
Ensure Reagent Supply Center covers are
securely in place.
Flush Water Lines option of the
START UP Procedure was not
performed.
Perform the Flush Water Lines option of the
START UP Procedure.
• Ensure water is present in the Reagent
Supply Center.
• If still unresolved, contact your Abbott
Representative.
Cuvette wash pump valve failure or
crimped tubing.
Contact your Abbott Representative.
System DI water source is not
functioning.
1. Check the DI water system.
2. Perform the Flush Water Lines option of the
START UP Procedure. Refer to
Section 9,
Service and Maintenance
of the
AEROSET
System Operations Manual
.
3. Check for water delivery to the Reagent
Supply Centers.
Bellows pump failure/valve failure.
Contact your Abbott Representative.
Refrigerator condensation pan contains bluish water
Probable Cause(s)
Corrective Action(s)
A bluish tint in the water originating
from the Reagent Supply Centers’
refrigerated condensation units is
caused by a coating within the unit.
This observation is normal. The
length of time of this situation varies,
relative to the daily amount of
condensation generated by the
instrument. Under normal
conditions, the bluish water should
not persist longer than one month
post-installation.
None required.