PROPRIETARY INFORMATION
CONFIDENTIAL MATERIAL
Not for Customer Distribution
Hardware
AEROSET
®
System Troubleshooting Guide
243
94816-107—November 2004
4. Verify the R1A probe is centered over the wash cup.
5.
6. Select <
Adjust
> on the right side of the screen.
If...
Then
R1A probe is not centered
over the wash cup
Proceed to step 6.
R1A probe is centered over
the wash cup
Proceed to step 9.