PROPRIETARY INFORMATION
CONFIDENTIAL MATERIAL
Software
Not for Customer Distribution
212
AEROSET
®
System Troubleshooting Guide
94816-107—November 2004
System Configuration
FAC Limit % Field—Configuration
The FAC Limit % field was added to the
Calibration
page of
the
Assay Configuration
screen with software version
1.02ER000. Acceptable entries in the FAC Limit % field range
from 0 to 99. After installation of software v1.02ER000, the
default setting is
10
.
If set to zero, the FAC Limit check is not
performed
.
For those assays configured on the AEROSET System at the
time of software installation, the default setting for the FAC
Limit % field is
10
.
Any assays imported (from the Import Disks listed below) after
upgrade to software v1.02ER000 will have a FAC Limit %
default setting of
0
and must be edited to 10.
The FAC Limit % was defined as 10 (%) in all versions of
AEROSET software prior to v1.02ER000, but this was not an
editable field. This value was used in validation testing for all
currently released Abbott assays and was applied to any
non-Abbott assays configured on the system prior to
v1.02ER000.
For non-Abbott assays or manually configured assays—the
default setting will be
10
for the FAC Limit % field.
Refer to
Assay Configuration
in
Section 2, Installation
Procedures and Special Requirements
of the
AEROSET System
Operations Manual
for instructions on editing the FAC
Limit % field.
• 4E43-02/4E43-05
Import Disk 1–50/Import Disk >50
• 4E43-08/4E43-09
Import Disk Drugs of Abuse v5.00
• 4E43-03/4E43-06
Import Disk Serum Proteins v3.0
• 4E43-04/4E43-07
Import Disk TDMs v4.00