94816-107—November 2004
DBil
(LN 7D59)
Bilirubin
Biliverdin
CAP survey results
show a 38.2% CV for
16 different chemistry
methodologies. The
AEROSET System
produces higher
biliverdin values than
most other analyzers.
A factor may be
entered to adjust. Not
intended for use with
neonatal specimens.
Abbott reagent
LN 8G63-20 is the
same reagent
formulation as DCL
labeled product
LN 2K59-20.
DIBIL
(LN
2K59-20,
DCL
Label)
Direct Bil 2,4-Dichlorophenyl diazonium salt
Azobilirubin
The formation of
azobilirubin is
measured
spectrophoto-
metrically and is
directly proportional
to the concentration
of bilirubin.
Avoid hemolysis
during sample
collection. Protect
samples from light.
Same formulation as
LN 8G63-20.
Dig*
D Antibody to Digoxin/BIS-TRIS buffer (R1) + Digoxin-coated
Microparticles (R2)
Digoxin:Antibody c Digoxin-coated Microparticle:Antibody
complex
Digoxin
concentration is
inversely proportional
to absorbance change.
(High absorbance
change = low digoxin
concentration; low
absorbance change =
high digoxin
concentration.)
Assay
Name
Reaction
Tips
* MULTIGENT
®
assay
Oxidation
Sodium Nitrite
Acid
Competition