![background image](http://html1.mh-extra.com/html/abbott/aeroset/aeroset_troubleshooting-manual_450971263.webp)
PROPRIETARY INFORMATION
CONFIDENTIAL MATERIAL
Not for Customer Distribution
Flowcharts
AEROSET
®
System Troubleshooting Guide
263
94816-107—November 2004
LIS/Communication Problems, Continued
4.
Host icon
with green line
through it (or not
BLUE)?
Host cable
connected?
Connect
cable
Cycle
power
Verify
configuration
matches on Host &
AEROSET
Changes
required?
Make changes
& cycle power
Run.
Resolved?
Yes
Run debug
Change to
AEROSET
configuration
required?
Yes
Cycle
power
No
No
Yes
Contact LIS
Vendor
Evaluate debug;
consult LIS
support
Yes
5.
No
Yes
No
No
Stop
Service
No
Yes
Run.
Resolved?
Run.
Resolved?
No
Host
suspended?
Yes
No
Yes
"Resume" Host
communiction