PROPRIETARY INFORMATION
CONFIDENTIAL MATERIAL
Not for Customer Distribution
Flowcharts
AEROSET
®
System Troubleshooting Guide
247
94816-107—November 2004
Abbott Assay—Basic Assay Troubleshooting
Customer
perceive assay
is performing
as expected?
Can
performance
change be traced
to a certain
event or date?
Yes
Text for assay
on ASSAYS
screen green?
Yes
Stop
Investigate the
following
bullet points
(next page)
Troubleshoot
event, or
investigate
implementation
of different lots/
shipments of
reagent or
calibrator for
date in question
Yes
No
No
No
Yes
No
Proceed to
QC OOR
Flowchart
QC OOR?
Proceed to
Illegal
Configuration or
Calibration
Errors
Flowchart