PROPRIETARY INFORMATION
CONFIDENTIAL MATERIAL
Hardware
Not for Customer Distribution
242
AEROSET
®
System Troubleshooting Guide
94816-107—November 2004
R1A Probe Wash Cup Alignment
1. Select the
Maintenance Utilities
icon in the Action Area
of the Main Display. The
MAINTENANCE UTILITIES
screen displays.
2. Select the
Rgt Area 1
tab. The
Rgt Area 1
page displays.
Figure 21: MAINTENANCE UTILITIES Screen, Rgt Area 1
Page
3. Select
Rgt Arm A WCup <Adj>
.
NOTE:
The reagent arm may not move if it is
centered over the wash cup.
Rgt Arm A
WCup <Adj>