![background image](http://html1.mh-extra.com/html/abbott/aeroset/aeroset_troubleshooting-manual_450971213.webp)
PROPRIETARY INFORMATION
CONFIDENTIAL MATERIAL
Not for Customer Distribution
Software
AEROSET
®
System Troubleshooting Guide
213
94816-107—November 2004
System Configuration Software Disk v3.0
The AEROSET System Configuration Software Disk v3.0 can
only be used with AEROSET software v1.00ER005.0 or
v1.00ER005.2.
Do not use System Configuration Software Disk v3.0 with
AEROSET software v1.02ER000 or greater. Installation of the
System Configuration Software Disk v3.0 on a v1.02ER000
AEROSET System adversely affects system parameters on the
Unit Configuration
and automatic
Startup/Shutdown
Setting
screens.
NOTE:
AEROSET System software version 1.03ER000
prevents restoration of System Configuration
parameters using a System Configuration disk created
from a different version of system software.
The
Unit Configuration
and automatic
Startup/Shutdown
Setting
screens should only be accessed by Abbott personnel.
Perform the following steps to access these screens.
1. Log On with the “extra” or “guest” level user code and
password.
2. Perform the steps below to access the
Unit Configuration
or automatic
Startup/Shutdown Setting
screens.
a. Press
[Alt]
+
[Print Screen]
on the keyboard at the
same time. The
CONTROL BOX
dialog window
displays.
b. Press
[
]
on the keyboard to highlight the Special
option then press
[Enter]
to select. A DOS window
with the c:\prompt displays.
c. Type the appropriate entry to access the desired
screen:
•
Type
unitconf
then press
[Enter]
to access the
Unit Configuration
screen.
•
Type
setups
then press
[Enter]
to access the
automatic
Startup/Shutdown Setting
screen.
3. Edit the appropriate fields back to the default settings
listed in the following sections. These screens should
match exactly, except those fields noted as customer
defined.
4. Select
<OK>
to save the settings and exit.