Hardware
AEROSET
®
System Troubleshooting Guide
235
94816-107—November 2004
4. Select
A-Line (Outer)
A <Adj
> to position the empty
cartridge in the R1A adjustment position.
5. Select
Rgt Arm A Horz Rgt <Adj>
.
CAUTION:
The R1A reagent probe moves over the
Reagent Carousel.
6. Verify the reagent probe is centered over the cartridge in
position A1.
7.
8. Select
<Adjust>
on the right side of the screen.
If...
Then
Reagent probe is not
centered over the cartridge
Proceed to step 8.
Reagent probe is centered
over the cartridge
Proceed to step 16.
!