PROPRIETARY INFORMATION
CONFIDENTIAL MATERIAL
Assays/Reagents
Not for Customer Distribution
160
AEROSET
®
System Troubleshooting Guide
94816-107—November 2004
Propoxyphene
Cross Reactivity Testing Performed on the AEROSET System by SYVA
Compounds listed in the following table yielded a negative
result at 300 ng/mL propoxyphene (Multi DOA Cutoff
calibrator) when tested at the concentration listed.
Compound
SYVA Package Insert
µg/mL
Acetaminophen
1,000
N
-Acetylprocainamide (NAPA)
400
Acetylsalicylic acid
1,000
Alpha-acetyl-
N
,
N
-dinormethadol
(dinor LAAM)
25
l
-Alpha-acetylmethodol (LAAM)
25
Buprenorphine
100
Cimetidine
1,000
Clomipramine
2.5
Clonidine
1,000
Cotinine
100
Diphenhydramine
1,000
Doxepin
10
Fluoxetine
500
Glutethimide
500
Ibuprofen
1,000
Ketamine
100
Ketorolac tromethamine
1,000
Lysergic acid diethylamide (LSD)
10 ng/mL
Meperidine
1,000
Naproxen
1,000
Phenytoin
1,000
Ranitidine
900
Scopolamine
500
Thioridazine
100
Tramadol
100