Hardware
AEROSET
®
System Troubleshooting Guide
231
94816-107—November 2004
Specifications
Water
USP Purified and WFI Requirements
Purified Water
Water for Injection
Conductivity
Not greater than 2.1 mS/cm at
25°C*
Not greater than 2.1 mS/cm at
25°C*
pH
5.0 to 7.0
5.0 to 7.0
Total Organic Carbon
Not greater than 500 ppb
Not greater than 500 ppb
Total Bioburden
Does not exceed
10,000 CFU/100 mL
Does not exceed
10,000 CFU/100 mL
Endotoxin
N/A
Not more than 0.25 Endotoxin
Unit per mL
Coliform Level
Zero per 100 mL
Zero per 100 mL
* Refer to USP 24/NF19 Monograph <645> “Water Conductivity” if conductivity meter is not
temperature compensated, or conductivity value exceeds 2.1 mS/cm at 25°C.
NOTE:
The above USP specifications are derived from the ASTM Type III Specification.