PROPRIETARY INFORMATION
CONFIDENTIAL MATERIAL
Not for Customer Distribution
Assays/Reagents
AEROSET
®
System Troubleshooting Guide
147
94816-107—November 2004
Reagent container bar code label not read
Probable Cause(s)
Corrective Action(s)
Bar code label not positioned
correctly.
Check positioning of the R1 and R2 reagent
bottles; verify the bar code label is facing the
center of the carousel.