Flowcharts
Not for Customer Distribution
252
AEROSET
®
System Troubleshooting Guide
94816-107—November 2004
Confirmation of Assay Parameter Settings (Illegal
Configuration), Continued
If everything is OK, and assay is still illegally configured (assay
text still black), verify the following:
1. Assay name is in range 1 to 9999.
2. If a calibrator is defined, it must be placed in the specified
position.
3. If a Use Factor calibration method is specified, the assay
for the referred calibration curve must be configured.
4. If a diluent is defined, it must be placed in the specified
position.
5. Sample Interference Indices—photometric assays to be
run must be specified and configured.
6. Calculation Assays—all assays used for the calculation
must be configured.
7. If a non-self (user defined assay file) blank is defined, the
specified test must be configured.
8. Multiple Rgt Ctgs option is ON and R2 and/or R2 reagent
is empty.
2