System Troubleshooting Guide
179
94816-107—November 2004
QC Statistics—Decimal Places
The decimal places used for QC statistics are based on the
value configured in the Decimal Places field on the
Base
page
of the
Assay Configuration
screen.
Mean = Decimal places value + 1 additional decimal place
SD
= Decimal places value + 1 additional decimal place
CV
= 2 Decimal places (fixed)
Range = Decimal places value
Decimal Places
Value (Base Page
of Assay
Configuration
Screen)
Decimal Places
of Mean Values
Decimal Places
of SD Values
Decimal Places
of CV Values
Decimal Places
of Range Values
1
2
2
2
1
2
3
3
2
2
3
4
4
2