PROPRIETARY INFORMATION
CONFIDENTIAL MATERIAL
Not for Customer Distribution
Assays/Reagents
AEROSET
®
System Troubleshooting Guide
115
94816-107—November 2004
Cannabinoids
Cross Reactivity Testing Performed on the AEROSET System by SYVA
Compounds listed in the following table yielded a negative
result at 50 ng/mL 11-nor-
∆
9
-THC-9-carboxylic acid (Multi
DOA Cutoff calibrator) when tested at the concentration
listed.
Compound
SYVA Package Insert
µg/mL
Acetaminophen
1,000
N
-Acetylprocainamide (NAPA)
400
Alpha-acetyl-
N
,
N
-dinormethadol
(dinor LAAM)
25
l
-Alpha-acetylmethodol (LAAM)
25
Buprenorphine
100
Caffeine
1,000
Cimetidine
1,000
Clomipramine
2.5
Clonidine
1,000
Codeine
500
Cotinine
100
Cyclobenzaprine
28
Desipramine
800
Diphenhydramine
1,000
Doxepin
10
2-Ethylidene-1,5-dimethyl-3,
3-diphenylpyrrolidine (EDDP)
1,000
Fluoxetine
500
Glutethimide
300
Ibuprofen
1,000
Ketamine
100
Ketorolac tromethamine
1,000
Lysergic acid diethylamide (LSD)
10 ng/mL
Naproxen
1,000
Nortriptyline
750
Phenytoin
1,000