Calculated Assay Results
Recalculating Calculated Assay Results
The assays included in a calculated assay
must
be performed
during the same run for the calculation to take place. The
result is calculated with the original assay results and does not
automatically recalculate with an assay’s rerun data.
If a retest is necessary for one of the assays of a calculated
result, retest all the assays included in the calculated result.
Use the following instructions to move an assay’s rerun data to
the original data, and to perform recalculation of the ratio.
1. On the
DATABASE
screen, select the rerun sample (rerun
sample has the Processing Code “L”).
2. Select
<Result>
in the right column of the
DATABASE
screen. The
RESULT
screen displays.
3. Display the desired rerun result.
4. Select
<Select>
in the right column of the
RESULT
screen.
5. Select
<OK>
on the
Select Result
dialog window. An “S”
displays to the left of the assay name and the rerun result
is copied over the original result for this assay.
6. Select
<OK>
to exit the
RESULT
screen.
7. On the
DATABASE
screen, select the original sample
result.
8. Select
<ReCalc>
in the right column. The
Recalc
dialog
window displays.
9. Select
<Select Assay>
then select the ratio to be
recalculated on the
SELECT ASSAYS
screen.
10. Select
<OK>
. The ratio is calculated from the rerun assay
data.