PROPRIETARY INFORMATION
CONFIDENTIAL MATERIAL
Assays/Reagents
Not for Customer Distribution
114
AEROSET
®
System Troubleshooting Guide
94816-107—November 2004
Calcium
Observed Problem
LOD/LOQ claim not duplicated by customer
Probable Cause(s)
Corrective Action(s)
Reagent cross contamination.
NOTE:
Creatinine
contaminates Calcium at very
low levels, even with
SmartWash™ configured. The
contamination is not clinically
significant.
Run Calcium in batch mode for LOD/LOQ
testing.