PROPRIETARY INFORMATION
CONFIDENTIAL MATERIAL
Flowcharts
Not for Customer Distribution
264
AEROSET
®
System Troubleshooting Guide
94816-107—November 2004
Water Issue
Stop
Service
1 Plug in
2 Turn on
3 Use another outlet
Light on?
Check external
water carboy;
water present?
Yes
Error Log Messages
99, 100, 240-248
Air in tubing?
Yes
Verify:
1 Cap plugged
in?
2 Switch on?
3 Outlet has
power?
Yes
1 Check tubing
connections &
crimps
2 Clean & replace
filter, if necessary
3 Cycle power to
Analyzer
4 Flush lines on
START UP screen
Service
External water
prior to
carboy?
Yes
Air still
present?
Carboy
overflow?
Check input
pressure: 30 to 40
psi desired. Adjust
if necessary
Pull up cap.
Does water
flow?
Yes
Yes
No
Check
external water
supply. OK?
Yes
Service
Proceed to
B
No
Flush water
lines. Resolved?
No
Yes
No
Yes
No
No
Yes
Contact
water vendor
Can customer
resolve?
No
No
Yes
No
1 Cycle power
2 Perform flush
water lines
Resolved?
A
B
Yes
Yes
Stop
No