Hardware
Not for Customer Distribution
240
AEROSET
®
System Troubleshooting Guide
94816-107—November 2004
4. Verify the R1A probe is centered over the A-Line cuvette in
the Reaction Carousel.
5.
6. Select <
Adjust
> on the right side of the screen.
If...
Then
R1A probe is not centered
over the cuvette
Proceed to step 6.
R1A probe is centered over
the cuvette
Proceed to step 9.
A-Line
Cuvette