![background image](http://html1.mh-extra.com/html/abbott/aeroset/aeroset_troubleshooting-manual_450971055.webp)
PROPRIETARY INFORMATION
CONFIDENTIAL MATERIAL
Not for Customer Distribution
Assays/Reagents
AEROSET
®
System Troubleshooting Guide
55
94816-107—November 2004
Prpx
Antibody to Propoxyphene (R1) + G-6-PD-Propoxyphene (R2) (Active)
G-6-PD-Propoxyphene:Antibody complex (Inactive)
G-6-PD-Propoxyphene:Antibody complex (Inactive) + Propoxyphene
G-6-PD-Propoxyphene (Active) + Propoxyphene:Antibody complex
G-6-PD-Propoxyphene (Active) + NAD + Glucose-6-phosphate
NADH + 6-phosphogluconate
G-6-PD = Glucose-6-phosphate dehydrogenase from bacterial source
NADH = Nicotinamide adenine dinucleotide
Use Multi DOA Cutoff
calibrator in duplicate
for calibration. Cutoff
calibrator value is
zero on
CALIBRATOR/
CONTROL
screen.
RF
RF (Sample) + Buffer (R1) + Antibody to RF adsorbed to latex particles (R2)
Turbidity due to agglutination of RF:Antibody:Particle complex
TBil
(LN 7D60)
Conjugated & Unconjugated Bilirubin
Biliverdin
Abbott reagent
LN 8G62-20 is the
same reagent
formulation as DCL
labeled product
LN 2K58-20.
Assay
Name
Reaction
Tips
* MULTIGENT
®
assay
Detergent
Solubilizer
Sodium Nitrite
Oxidation