Assays/Reagents
Not for Customer Distribution
38
AEROSET
®
System Troubleshooting Guide
94816-107—November 2004
*3-carboxy-4-nitroaniline
Use Cal Factor From
Use Cal Factor From option only displays if
Use Factor/Blank
is selected as the calibration mode. This method allows the
operator to enter the name of the assay from which the
calibration data are used.
This list includes all the assays configured on the system.
This method may be used when two assays use the same
reagent and have the same sample-to-reagent volume ratios.
Rather than calibrating both assays, only one of the assays is
calibrated and the other assay is configured to refer to this
calibration data for result calculation. For example, a urine
application of an assay may use the calibration data generated
for the serum application for the same assay,
e.g
., urine
Amylase.
The assay selected as the reference assay must generate its own
curve and cannot also be defined with the Use Factor/Blank
calibration mode.
Assay
Chromophore
Wavelength
Total
Volume
µL
Sample
Volume
µL
Extinction
Factor
Enzyme
Factor
Acid
Phosphatase
Alpha-
naphthol-Fast
Red TR
complex
412/660
162.0
12.0
12.9
1046.5
Alkaline
Phosphatase
PNP
404/476
257.0
7.0
17.08
2150
Amylase
CNP
404/476
204.8
4.8
12.44
3431
ALT
NADH
340/380
205.3
5.3
4.76
8141
ALT Activated
NADH
340/380
178.7
3.7
4.76
10151
AST
NADH
340/380
205.3
5.3
4.76
8141
AST Activated
NADH
340/380
182.4
7.4
4.76
5180
CK
NADH
340/412
204.0
4.0
5.68
8974
GGT
A PNA*
412/660
204.0
4.0
6.31
8077
LD
NADH
340/380
203.2
3.2
4.76
13341