PROPRIETARY INFORMATION
CONFIDENTIAL MATERIAL
Assays/Reagents
Not for Customer Distribution
44
AEROSET
®
System Troubleshooting Guide
94816-107—November 2004
AST-A
L
-Asp
α
-Ketoglutarate
Oxaloa
L
-glutamate
Oxaloa NADH
L
- NAD
+
NADH = Nicotinamide adenine dinucleotide
MD = Malate dehydrogenase
Activated reagent
contains pyridoxal-5-
phosphate (P-5-P).
Do not use plasma
samples collected
using ammonium
heparin.
Barb
Antibody to Barbiturates (R1) + G-6-PD-Barbiturates (R2) (Active)
G-6-PD-Barbiturates:Antibody complex (Inactive)
G-6-PD-Barbiturates:Antibody complex (Inactive) + Barbiturates
G-6-PD-Barbiturates (Active) + Barbiturates:Antibody complex
G-6-PD-Barbiturates (Active) + NAD + Glucose-6-phosphate
NADH + 6-phosphogluconate
G-6-PD = Glucose-6-phosphate dehydrogenase from bacterial source
NADH = Nicotinamide adenine dinucleotide
Use Multi DOA Cutoff
calibrator in duplicate
for calibration. Cutoff
calibrator value is
zero on
CALIBRATOR/
CONTROL
screen.
Benz
Antibody to Benzodiazepines (R1) + G-6-PD-Benzodiazepines (R2) (Active)
G-6-PD-Benzodiazepines:Antibody complex (Inactive)
G-6-PD-Benzodiazepines:Antibody complex (Inactive) + Benzodiazepines
G-6-PD-Benzodiazepines (Active) + Benzodiazepines:Antibody complex
G-6-PD-Benzodiazepines (Active) + NAD + Glucose-6-phosphate
NADH + 6-phosphogluconate
G-6-PD = Glucose-6-phosphate dehydrogenase from bacterial source
NADH = Nicotinamide adenine dinucleotide
Use Multi DOA Cutoff
calibrator in duplicate
for calibration. Cutoff
calibrator value is
zero on
CALIBRATOR/
CONTROL
screen.
Does not correlate
well with AxSYM
®
Benzodiazepines
assay.
Assay
Name
Reaction
Tips
* MULTIGENT
®
assay
AST, P-5-P
MD