94816-107—November 2004
BilD
(LN
8G63-20)
Direct (Conjugated) Bil 2,4-Dichlorophenyl diazonium salt
Azobilirubin
The formation of
azobilirubin is
measured
spectrophoto-
metrically and is
directly proportional
to the concentration
of bilirubin.
Avoid hemolysis
during sample
collection. Protect
samples from light.
Same formulation as
LN 2K59-20.
BilT
(LN
8G62-20)
Total (Conjugated & Unconjugated) Bil 2,4-Dichlorophenyl
diazonium salt
Azobilirubin
The formation of
azobilirubin is
measured
spectrophoto-
metrically and is
directly proportional
to the concentration
of bilirubin.
Avoid hemolysis
during sample
collection. Protect
samples from light.
Same formulation as
LN 2K58-20.
C3
Complement C3 + PEG/buffer (R1) + Antibody to C3 (R2)
Turbidity
due to C3:Antibody complex
PEG = Polyethylene glycol
C4
Complement C4 + PEG/buffer (R1) + Antibody to C4 (R2)
Turbidity
due to C4:Antibody complex
PEG = Polyethylene glycol
Analyze fresh or
refrigerate at 2 to 8°C
no more than two
days.
Ca
2 Arsenazo III + Ca
+2
Ca-Arsenazo complex
+2
(blue-purple)
Must not be on the
same line as the
Lipase assay.
New reference range:
9.0 to 10.4 mg/dL.
Assay
Name
Reaction
Tips
* MULTIGENT
®
assay
Acid
Surfactant