Hardware
AEROSET
®
System Troubleshooting Guide
233
94816-107—November 2004
7. Select
<Data>
. The
Photometer and ICT™ A/D Data
screen displays.
Figure 17: Photometer and ICT A/D Data Screen
8. Observe the displayed Gain and A/D values. Ensure the
Gain values are
울
2.6
AND
A/D readings are > 35,000,
with a range of < 300.
NOTE:
If the Gain value is 2.6
AND
A/D reading
is
울
35,000, or if the Gain value is 4.0, replace the
lamp. Refer to
Source Lamp Replacement
in
Section 9, Service and Maintenance
of the
AEROSET
System Operations Manual
. After lamp
replacement, recheck the Gain and A/D values;
verify all Gain values are
울
2.6 and A/D readings
are > 35,000. If not resolved, contact your Abbott
Representative.
9. Print the screen for reference. Select
<OK>
. The
Rxn Area
page of the
MAINTENANCE UTILITIES
screen displays.
10. Select
Shutter <OFF>
to close the shutter.
11. Using the Rxn Crsl buttons, rotate the Reaction Carousel
to replace the cuvette segment.
12. Select
Rxn Crsl <Home>
.
13. Select
<OK>
to exit the
Rxn Area
page.
14. Select
<Cancel>
.
Gain and A/D Values