Hardware
Not for Customer Distribution
238
AEROSET
®
System Troubleshooting Guide
94816-107—November 2004
6. Verify reagent probe alignment over the reagent cartridge.
7. Select
<Adjust>
on the right side of the screen.
8. Select <
CW
> (clockwise) or <
CCW
> (counterclockwise) as
needed to center the probe over the reagent cartridge.
9. Select
<OK>
to save the position.
10. Select
Rgt Arm A Vert Down <Adj>
.
11. Verify the reagent arm position.
12.
13. Select
Rgt Arm
A Vert Home/Up <Adj>
.
14. Repeat steps 4 through 12.
If...
Then
Reagent probe is not
aligned over the reagent
cartridge
Proceed to step 13.
Reagent probe is aligned
over the reagent
cartridge
Select
<OK>
on the
MAINTENANCE UTILITIES
screen then select
<OK>
on
the
Home All Robotics
dialog
window.
Counterclockwise
Clockwise