7
•
Maintenance and Service
Pumps C, Pumps F ( VC-Pxx, VF-Pxx)
Operating Manual
Page 161
10. For about 5 minutes, observe the liquid level in the tubing. Finish the
observation before a new seal wash cycle starts.
11. Depends on your observation:
Observation
Remedial Action
Liquid level in the tubing is
increasing again
1. Replace the piston seals in the
left
pump
head.
2. Reconnect the seal wash tubing to the right
pump head.
3. Reconnect the seal wash tubing to the seal
wash detector (port labeled
in
).
4. Repeat the test starting with step 1.
Liquid level in the tubing does not
increase again
1. Replace the piston seals in the
right
pump
head.
2. Reconnect the seal wash tubing to the right
pump head.
3. Reconnect the seal wash tubing to the seal
wash detector (port labeled
in
).
4. Repeat the test starting with step 1.
TIP
If you observe leakage from a pump head, verify that the seal wash
tubing is properly connected to the pump head. If the tubing is not
properly connected or if the tubing is crimped or kinked, seal wash liquid
may leak into the pump.
Содержание thermoscientific Vanquish Pumps C
Страница 4: ...Contacting Us Page 4 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 30: ...2 Safety Page 30 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 40: ...3 Device Overview Page 40 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 196: ...7 Maintenance and Service Page 196 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 214: ...8 Troubleshooting Page 214 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 284: ...12 Accessories Consumables and Replacement Parts Page 284 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 290: ...13 Appendix Page 290 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Страница 296: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...